FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1891372 · Received November 2, 2010

Report

Report Number
1824206-2010-10813
Event Type
Malfunction
Date Received
November 2, 2010
Date of Event
October 7, 2010
Report Date
October 7, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE PROBLEM TO BE A FAULTY EMERGENCY TREND VALVE. THE EMERGENCY TREND FUNCTION STILL WORKS. THE TECHNICIAN REPLACED THE EMERGENCY TREND VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD HI-LOW DRIFTS DOWN SLOWLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1