VASOVIEW HEMOPRO 2

MDR report key: 18913405 · Received March 15, 2024

Report

Report Number
2242352-2024-00250
Event Type
Malfunction
Date Received
March 15, 2024
Date of Event
February 24, 2024
Report Date
May 30, 2024
Manufacturer
MAQUET CARDIOVASCULAR LLC
Product Code
GEI
UDI-DI
00607567700406
PMA / PMN Number
K101274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED.

Additional Manufacturer Narrative · 0

TRACKWISE#: (B)(4). THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION ON 03/15/2024. A PHOTOGRAPH AND VIDEO WAS PROVIDED BY THE ACCOUNT. A PHOTOGRAPHIC EVALUATION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED. A VIDEO INSPECTION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED. A SYRINGE WITH FLUID WAS USED TO INFLATE THE BTT. THE BTT DID NOT INFLATE AND A SMALL STREAM OF LIQUID WAS OBSERVED COMING OUT OF THE BTT. NO OTHER VISUAL DEFECTS WERE OBSERVED. AN INVESTIGATION WAS CONDUCTED ON 03/19/2024. A VISUAL INSPECTION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED. A 50 CC SYRINGE, FILLED WITH AIR WAS ATTACHED TO THE INFLATION PORT LINE ON THE BTT AND SQUEEZED THE SYRINGE TO INJECT AIR. THE BTT WAS UNABLE TO BE INFLATED. A SYRINGE WITH FLUID WAS USED TO INFLATE THE BTT. THE BTT DID NOT INFLATE AND A SMALL STREAM OF LIQUID WAS OBSERVED COMING OUT OF THE BTT. A LEAK WAS OBSERVED ON THE SIDE OF THE SILICONE BALLOON. AN ENGINEER EVALUATION WAS CONDUCTED ON 07-MAY-2024. THE FOLLOWING IS MANUFACTURING ENGINEERING¿S EVALUATION OF THE VH-4000 HEMOPRO2, BTT SILICONE BALLOON SUBASSEMBLY ((B)(6)), ONETRACK COMPLAINT #(B)(4), THAT WAS RETURNED FOR THE REPORTED FAILURE OF ¿INFLATION ISSUE¿. EVALUATION DETAILS: THE COMPLAINT WAS CONFIRMED. THE FIRST TEST, FOR THE BTT SILICONE BALLOON SUBASSEMBLY, WAS TO INFLATE THE BALLOON WITH 25 CC OF AIR FROM THE COMPLAINT LAB¿S SYRINGE. THE AIR LEAKED OUT AND THE BTT SILICONE BALLOON DID NOT INFLATE AT ALL. TO DETERMINE THE LOCATION OF THE LEAK, A LEAK TEST IN A WATER BATH WAS PERFORMED. TWO (2) LEAKS ALONG THE PROXIMAL SUTURE GROOVE WERE FOUND. REFER TO FIGURE 1. SEE ATTACHED ENGINEER EVALUATION MEMO. BASED ON THE CONDITION OF THE DEVICE, THE REPORTED FAILURE "INFLATION ISSUE" WAS CONFIRMED.

Additional Manufacturer Narrative · 0

TRACKWISE#: (B)(4). UPDATED SECTIONS: B4, D9, G3, G6, H2, H3, H6, H10. THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION ON 03/15/2024. A PHOTOGRAPH AND VIDEO WAS PROVIDED BY THE ACCOUNT. A PHOTOGRAPHIC EVALUATION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED. A VIDEO INSPECTION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED. A SYRINGE WITH FLUID WAS USED TO INFLATE THE BTT. THE BTT DID NOT INFLATE AND A SMALL STREAM OF LIQUID WAS OBSERVED COMING OUT OF THE BTT. NO OTHER VISUAL DEFECTS WERE OBSERVED. AN INVESTIGATION WAS CONDUCTED ON 03/19/2024. A VISUAL INSPECTION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED. A MECHANICAL INVESTIGATION WAS CONDUCTED. A 5CC SYRINGE, FILLED WITH SALINE WAS ATTACHED TO THE CO2 LINE ON THE BTT AND SQUEEZED THE SYRINGE TO SPRAY THE SALINE. THERE WAS NO BACKFLOW OBSERVED IN THE CO2 LINE. A LEAK WAS OBSERVED ON THE SIDE OF THE SILICONE BALLOON. THE BTT WAS UNABLE TO BE INFLATED. NO VISUAL DEFECTS WERE OBSERVED ON THE SILICONE BTT. BASED ON THE CONDITION OF THE DEVICE, THE REPORTED FAILURE "INFLATION ISSUE" WAS CONFIRMED. SPECIFIC ACTIONS FOR THE REPORTED FAILURE MODE IS BEING MAINTAINED AND DOCUMENTED UNDER MAQUET'S CORRECTIVE AND PREVENTIVE ACTION (CAPA) SYSTEM. THE LOT # 3000345542 HISTORY RECORD REVIEW WAS COMPLETED. THERE WERE NO NCMRS, REWORK, OR DEVIATIONS DOCUMENTED FOR THE REPORTED LOT NUMBER. BASED ON THE DHR/LHR REVIEW RESULTS, IT WAS DETERMINED THAT THERE IS NO RELATION BETWEEN THE BATCH MANUFACTURING PROCESS AND THE REPORTED FAILURE.

Description of Event or Problem · 0

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIEW HEMOPRO 2 BTT PORT BALLOON FAILED TO MAINTAIN DESIRED INFLATION SIZE. PER VIDEO PROVIDED BY THE COMPLAINANT, THE BALLOON DID NOT INFLATE AND THERE WAS EVIDENCE OF BLOOD. UPON INSPECTION, A LEAKAGE WAS FOUND AT THE RIM OF THE BALLOON, LEADING TO ITS INABILITY TO MAINTAIN THE DESIRED INFLATION SIZE. THEY PROCEEDED WITH THE PROCEDURE DUE TO ADEQUATE TUNNEL VISUALIZATION. NO DELAY. THE PATIENT WAS UNHARMED AND IS IN STABLE CONDITION.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
535329 VASOVIEW HEMOPRO 2 ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI MAQUET CARDIOVASCULAR LLC VH-4000 3000345542 00607567700406

Patients

Seq Age Sex Outcome Treatment
1 71 YR Prefer Not To Disclose