PROXIMATE*ILS CURVED CIR STAPL
Report
- Report Number
- 3005075853-2010-06296
- Event Type
- Malfunction
- Date Received
- November 4, 2010
- Date of Event
- October 15, 2010
- Report Date
- October 21, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). WASHER UNCUT THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION WITHOUT STAPLES PRESENT AND WITH THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IN ADDITION, FIVE STAPLES PARTIALLY FORMED WERE RECEIVED INSIDE A SMALL BAG. IT SHOULD BE NOTED THAT IF THE INDICATOR IS NOT FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE OR IF THE FIRING HANDLE IS NOT FIRMLY SQUEEZED USING STEADY PRESSURE UNTIL THE FIRING HANDLE IS FULLY PARALLEL TO THE INSTRUMENT HANDLE, STAPLES COULD BE DEPLOYED WITHOUT COMPLETELY FORMING AND WITHOUT CUTTING THE WASHER. FOR MORE INFORMATION PLEASE REFER TO THE INSTRUCTIONS FOR USE. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4) : INFORMATION WAS NOT PROVIDED BY THE AFFILIATE. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. (B)(6).
IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, THE DEVICE WAS USED TO ANASTOMOSE THE COLON AND THE RECTUM. THE PUNCHED OUT SOUND OF THE WASHER WAS NOT HEARD AND THE DEVICE COULD NOT BE ANASTOMOSED PROPERLY AT THE FIRST FIRING. THE RECTUM WAS ANASTOMOSED AGAIN AND THE OPERATION WAS FINISHED WITHOUT ANY PROBLEM. REINFORCEMENT PRODUCT WAS NOT USED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE*ILS CURVED CIR STAPL | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |