FDA Adverse Event Malfunction Summary report: N

ACCURUS 800CS

MDR report key: 1891043 · Received October 29, 2010

Report

Report Number
2028159-2010-02092
Event Type
Malfunction
Date Received
October 29, 2010
Date of Event
October 1, 2010
Report Date
October 1, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K911808
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY SERVICE REP EXAMINED THE SYSTEM AND CONFIRMED THE SYSTEM MESSAGE REPORTED. THE COMMUNICATION CABLES WERE RESEATED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE NURSE REPORTED A SYSTEM MESSAGE DISPLAYED DURING THE CASE. THE SYSTEM WAS SWITCHED OUT AND THE CASE WAS COMPLETED AS PLANNED. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCURUS 800CS PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER ACCURUS NA

Patients

Seq Age Sex Outcome Treatment
1 UNK