FDA Adverse Event Malfunction Summary report: N

MINI TREK CORONARY DILATATION CATHETER

MDR report key: 1891006 · Received October 29, 2010

Report

Report Number
2024168-2010-02303
Event Type
Malfunction
Date Received
October 29, 2010
Date of Event
October 6, 2010
Report Date
October 6, 2010
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
LOX
PMA / PMN Number
P810046
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BALLOON MATERIAL RUPTURES CAN BE AFFECTED BY NUMEROUS FACTORS INCLUDING, BUT NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, HANDLING, LESION CHARACTERISTICS, PT DISEASE STATE, AN INTERACTION WITH A PREVIOUSLY IMPLANTED STENT, INSUFFICIENT PREPARATION PRIOR TO USE OR FROM INTERACTIONS WITH OTHER DEVICES. SINCE THERE WAS NO REPORT OF ANY LEAKS NOTED DURING PREPARATION FOR USE, THIS MAY INDICATE THAT THE BALLOON WAS NOT DAMAGED PRIOR TO THE PROCEDURE. ADDITIONALLY, THE LESION WAS REPORTED AS MODERATELY CALCIFIED AND MAY HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY. IF THE BALLOON EXPERIENCES MECHANICAL DAMAGE AT SOME POINT DURING THE PROCEDURE, THIS CAN CAUSE THE MATERIAL TO WEAKEN AND RUPTURE DURING AN ATTEMPT TO INFLATE THE CATHETER. ULTIMATELY, RETURN OF THE MINI TREK MAY HAVE AIDED THE EVALUATION IN DETERMINING A CAUSE FOR THE RUPTURE. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT AND ALL LOT RELEASE TESTING MET MANUFACTURING CRITERIA. THERE DOES NOT APPEAR TO BE ANY PRODUCT QUALITY DEFICIENCIES ASSOCIATED WITH THE LOT. BASED ON THE INFORMATION REC'D WITH THIS COMPLAINT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED BALLOON RUPTURE CANNOT BE DETERMINED. ALL DILATATION CATHETERS ARE 100% VISUALLY INSPECTED AND LEAK TESTED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A SAMPLING OF UNITS IS DESTRUCTIVELY TESTED TO VERIFY RBP AND BALLOON INTEGRITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE OBTUSE MARGINAL (OM). THE MINI TREK BALLOON RUPTURED AT 12 ATMOSPHERE (ATM). THERE WAS NO REPORTED PT EFFECT. THE PROCEDURE WAS COMPLETED USING A NON-ABBOTT DILATATION CATHETER AT 12 ATM. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINI TREK CORONARY DILATATION CATHETER LOX ABBOTT VASCULAR-CARDIAC THERAPIES NA 0070261

Patients

Seq Age Sex Outcome Treatment
1 UNK DILATATION CATHETER: MAVERICK BALLOON