FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 1890643
·
Received October 25, 2010
Report
- Report Number
- 1811755-2010-01497
- Event Type
- Malfunction
- Date Received
- October 25, 2010
- Date of Event
- September 15, 2010
- Report Date
- October 5, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH THE MOTOR AND BEARINGS, WHICH WERE EACH REPLACED ALONG WITH OTHER COMPONENTS. SERVICE DID A CLEAN, LUBE, AND ADJUST TO THE DEVICE, AND IT WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING AN ORAL SURGERY. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |