FDA Adverse Event
Malfunction
Summary report: N
SECURE 3 BED/SURG BED
MDR report key: 1890453
·
Received October 27, 2010
Report
- Report Number
- 1831750-2010-03295
- Event Type
- Malfunction
- Date Received
- October 27, 2010
- Date of Event
- September 28, 2010
- Report Date
- September 28, 2010
- Manufacturer
- STRYKER CORP., MEDICAL DIV.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
I-BED SYSTEM. MANUFACTURERS INVESTIGATION IS STILL ONGOING. IF ANY ADDITIONAL INFO IS RECEIVED, A F/U WILL BE ISSUED.
Description of Event or Problem · 1
IT HAS BEEN REPORTED THAT THE IBED AWARENESS LIGHTS ARE NOT WORKING. NO ADVERSE CONSEQUENCES HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SECURE 3 BED/SURG BED | A/C POWERED HOSPITAL BED | FNL | STRYKER CORP., MEDICAL DIV. | 3002S3EX | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |