FDA Adverse Event Malfunction Summary report: N

BD PHOENIX PANEL NMIC-306

MDR report key: 18903842 · Received March 14, 2024

Report

Report Number
1119779-2024-00205
Event Type
Malfunction
Date Received
March 14, 2024
Date of Event
February 2, 2024
Report Date
August 2, 2024
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
00382904492922
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY THIS COMPLAINT IS FOR HIGH MIC ERTAPENEM (ETP) WITH ENTEROBACTER CLOACAE WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER 3311874. THE CUSTOMER PROVIDED PHOENIX GENERATED LAB REPORTS AND AN ISOLATE BUT DID NOT PROVIDE PANEL RETURNS FOR THE INVESTIGATION. THE LAB REPORTS SHOW A RESISTANT RESULT FOR ETP WITH E. CLOACAE WHEN USING THE COMPLAINT BATCH. TO INVESTIGATE, THREE RETENTION PANELS OF THE COMPLAINT BATCH WERE INOCULATED WITH CUSTOMER RETURNED ISOLATE E. CLOACAE AND PLACED IN A PHOENIX M50 FOR ETP MIC RESULTS. IN ADDITION, ONE CONTROL PANEL FROM THE SAME MATERIAL BUT DIFFERENT BATCH WAS INOCULATED WITH CUSTOMER RETURNED ISOLATE E. CLOACAE AND PLACED IN A PHOENIX M50 FOR ETP MIC RESULTS. LAST, KIRBY-BAUER TESTING WAS PERFORMED ON THE CUSTOMER RETURNED ISOLATE WITH ETP. RESULTS OF THE TEST SHOW ALL PANELS RETURNED THE EXPECTED RESULTS FOR ETP ACCORDING TO DISC DIFFUSION RESULTS. THIS COMPLAINT IS NOT CONFIRMED. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. A REVIEW OF COMPLAINTS REVEALED NO OTHER COMPLAINTS ON THIS BATCH. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING WITH BD PHOENIX PANEL NMIC-306, ENTEROBACTER CLOACAE COMPLEX WAS FALSELY RESISTANT TO ERTAPENEM. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING WITH BD PHOENIX PANEL NMIC-306, ENTEROBACTER CLOACAE COMPLEX WAS FALSELY RESISTANT TO ERTAPENEM. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2076729 BD PHOENIX PANEL NMIC-306 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY LON BECTON DICKINSON & CO. (SPARKS) 3311874 00382904492922

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown