BD PHOENIX PANEL NMIC-306
Report
- Report Number
- 1119779-2024-00205
- Event Type
- Malfunction
- Date Received
- March 14, 2024
- Date of Event
- February 2, 2024
- Report Date
- August 2, 2024
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 00382904492922
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY THIS COMPLAINT IS FOR HIGH MIC ERTAPENEM (ETP) WITH ENTEROBACTER CLOACAE WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER 3311874. THE CUSTOMER PROVIDED PHOENIX GENERATED LAB REPORTS AND AN ISOLATE BUT DID NOT PROVIDE PANEL RETURNS FOR THE INVESTIGATION. THE LAB REPORTS SHOW A RESISTANT RESULT FOR ETP WITH E. CLOACAE WHEN USING THE COMPLAINT BATCH. TO INVESTIGATE, THREE RETENTION PANELS OF THE COMPLAINT BATCH WERE INOCULATED WITH CUSTOMER RETURNED ISOLATE E. CLOACAE AND PLACED IN A PHOENIX M50 FOR ETP MIC RESULTS. IN ADDITION, ONE CONTROL PANEL FROM THE SAME MATERIAL BUT DIFFERENT BATCH WAS INOCULATED WITH CUSTOMER RETURNED ISOLATE E. CLOACAE AND PLACED IN A PHOENIX M50 FOR ETP MIC RESULTS. LAST, KIRBY-BAUER TESTING WAS PERFORMED ON THE CUSTOMER RETURNED ISOLATE WITH ETP. RESULTS OF THE TEST SHOW ALL PANELS RETURNED THE EXPECTED RESULTS FOR ETP ACCORDING TO DISC DIFFUSION RESULTS. THIS COMPLAINT IS NOT CONFIRMED. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. A REVIEW OF COMPLAINTS REVEALED NO OTHER COMPLAINTS ON THIS BATCH. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY.
IT WAS REPORTED WHILE TESTING WITH BD PHOENIX PANEL NMIC-306, ENTEROBACTER CLOACAE COMPLEX WAS FALSELY RESISTANT TO ERTAPENEM. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
IT WAS REPORTED WHILE TESTING WITH BD PHOENIX PANEL NMIC-306, ENTEROBACTER CLOACAE COMPLEX WAS FALSELY RESISTANT TO ERTAPENEM. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2076729 | BD PHOENIX PANEL NMIC-306 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY | LON | BECTON DICKINSON & CO. (SPARKS) | 3311874 | 00382904492922 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |