DA VINCI S SURGICAL SYSTEM
Report
- Report Number
- 2955842-2010-00479
- Event Type
- Other
- Date Received
- November 3, 2010
- Date of Event
- October 4, 2010
- Report Date
- October 4, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081207
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING (FSE) FOUND SYSTEM ERROR CODE 23017 TO BE ASSOCIATED WITH A PATIENT SIDE MANIPULATOR (PSM) ARM. THE PSM IS AN INSTRUMENT ARM LOCATED ON THE PATIENT SIDE CART AND PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED PSM. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PATIENT. SYSTEM ERROR CODE 23017 APPEARS WHEN THE DA VINCI S SAFETY SYSTEM DETERMINES THAT A MOTOR DID NOT RESPOND AS EXPECTED AND THE MEASURED MOTION DID NOT MATCH THE INTERNAL SIMULATION OF THE MOTOR. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEMS PUT DA VINCI IN A RECOVERABLE SAFE STATE. THE PSM WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING CONFIRMED THE CUSTOMER REPORTED PROBLEM AND FOUND THE PSM TO HAVE A DEFECTIVE MOTOR/ENCODER ASSEMBLY WHICH WAS REPLACED TO REPAIR THE PSM. AS OF (B)(6) 2010, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING THE BEGINNING OF A DA VINCI S PROSTATECTOMY PROCEDURE, THE SURGICAL STAFF EXPERIENCED SYSTEM ERROR CODE 23017 ON A PATIENT SIDE MANIPULATOR (PSM) ARM. THE SITE RESTARTED THE SYSTEM MULTIPLE TIMES WITHOUT SUCCESS. AT THIS TIME, THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED PROCEDURE. NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI S SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS2000 A5.1P8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI S SYSTEM INSTRUMENT, ACCESSORIES AND ESU |