FDA Adverse Event
Malfunction
Summary report: N
IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM
MDR report key: 1890377
·
Received November 3, 2010
Report
- Report Number
- 2050012-2010-01146
- Event Type
- Malfunction
- Date Received
- November 3, 2010
- Date of Event
- October 12, 2010
- Report Date
- November 3, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K962294
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) VISITED THE SITE ON (B)(4) 2010. THE CUSTOMER INFORMED THE FSE THAT THE SYSTEM HAD A PLUG IN THE SAMPLE WASH CUP EARLIER THE DAY AND IT WAS FLUSHED OUT WITH BLEACH. THE FSE REPLACED THE CUVETTE WASH LINE FILTER AS A PRECAUTION. THE FSE FLUSHED THE LINE AND WASHED ALL CUVETTES. THE SYSTEM IS NOW OPERATING ACCEPTABLY.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO DRIED WASH SOLUTION OBSERVED UNDER IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO OPERATOR EXPOSURE WAS NOTED. THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM | CLINICAL IMMUNOCHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | IMMAGE 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |