FDA Adverse Event Malfunction Summary report: N

IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM

MDR report key: 1890377 · Received November 3, 2010

Report

Report Number
2050012-2010-01146
Event Type
Malfunction
Date Received
November 3, 2010
Date of Event
October 12, 2010
Report Date
November 3, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K962294
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BCI FIELD SERVICE ENGINEER (FSE) VISITED THE SITE ON (B)(4) 2010. THE CUSTOMER INFORMED THE FSE THAT THE SYSTEM HAD A PLUG IN THE SAMPLE WASH CUP EARLIER THE DAY AND IT WAS FLUSHED OUT WITH BLEACH. THE FSE REPLACED THE CUVETTE WASH LINE FILTER AS A PRECAUTION. THE FSE FLUSHED THE LINE AND WASHED ALL CUVETTES. THE SYSTEM IS NOW OPERATING ACCEPTABLY.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO DRIED WASH SOLUTION OBSERVED UNDER IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO OPERATOR EXPOSURE WAS NOTED. THERE WAS NO EFFECT TO PATIENT OR USER.