FDA Adverse Event Malfunction Summary report: N

VOYAGER NC CORONARY DILATATION CATHETER

MDR report key: 1890301 · Received October 22, 2010

Report

Report Number
2024168-2010-02231
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 27, 2010
Report Date
September 28, 2010
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
LOX
PMA / PMN Number
P810046
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). FACTORS THAT CAN CONTRIBUTE TO BALLOON RUPTURES INCLUDE, BUT NOT LIMITED TO, BALLOON DAMAGE DURING MFG, WEAK MATERIALS, EXCESSIVELY APPLIED PRESSURE, OR INTERACTIONS WITH ACCESSORY DEVICES, PT ANATOMY, AND/OR LESION CALCIFICATION OR TORTUOSITY. THE PT ANATOMY WAS MILDLY TORTUOUS AND CALCIFIED AND THE DEVICE WAS USED TO POST DILATE A STENT, WHICH LIKELY CONTRIBUTED TO THE RUPTURE. IT IS POSSIBLE THE BALLOON MATERIAL WAS DAMAGED DURING INTERACTIONS WITH OTHER DEVICES, OR PT ANATOMY, SUCH THAT THE BALLOON RUPTURED UPON INFLATION. REVIEW OF THE DEVICE HISTORY RECORD DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT. THE DEVICE WAS NOT RETURNED; THEREFORE, NO CONCLUSIVE CAUSE COULD BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LESION WAS LOCATED IN THE DISTAL RIGHT CORONARY ARTERY (RCA), WHICH WAS MILDLY TORTUOUS, HEAVILY CALCIFIED AND HAD A 90% STENOSIS. A ROTABLATOR WAS USED IN THE LESION AND A XIENCE V WAS IMPLANTED. THEN, THE VOYAGER NC BALLOON CATHETER WAS USED FOR POST-DILATATION, BUT THE BALLOON RUPTURED WHEN INFLATED FOR THE SECOND TIME. THEN, INTRAVASCULAR ULTRASOUND (IVUS) WAS PERFORMED AND THE STENT WAS CONFIRMED TO BE ADEQUATELY EXPANDED. REPORTEDLY, THERE WERE NO PT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL EVENT OR PT INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VOYAGER NC CORONARY DILATATION CATHETER LOX ABBOTT VASCULAR-CARDIAC THERAPIES NA 9050461

Patients

Seq Age Sex Outcome Treatment
1 DIL CATH: VOYAGER| STENT: XIENCE V| GUIDE WIRE: RUNTHROUGH| GUIDE CATH: FINECROSS| GUIDE CATH: ROTA