FDA Adverse Event Malfunction Summary report: N

REALIZE ADJ GASTRIC BAND-C

MDR report key: 1890212 · Received November 3, 2010

Report

Report Number
3005992282-2010-00340
Event Type
Malfunction
Date Received
November 3, 2010
Date of Event
October 18, 2010
Report Date
October 18, 2010
Manufacturer
OBTECH MEDICAL SARL_
Product Code
LTI
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DISPLACED. THE BAND/BALLOON WITH 50CM OF CATHETER, THE INJECTION PORT WITH THE LOCKING CONNECTOR AND THE TUBING STRAIN RELIEF WERE RETURNED FOR ANALYSIS. UPON VISUAL INSPECTION, BIOLOGICAL DEBRIS WAS OBSERVED AT ORIGINAL LOCATION OF THE TUBING STRAIN RELIEF. THEREFORE MIGRATION OF THE TUBING STRAIN RELIEF CAN BE CONFIRMED. THE STRAIN RELIEF WAS FLOATING ON THE CATHETER AT 42CM FROM THE LOCKING CONNECTOR. NO MARK OR DAMAGE WAS OBSERVED ON THE TUBING STRAIN RELIEF. HOWEVER, NO LINK CAN BE ESTABLISHED BETWEEN MOVEMENT OF THE STRAIN RELIEF AND LACK OF WEIGHT LOSS OF THE PATIENT. TO MITIGATE THE STRAIN RELIEF "MIGRATION" FROM THE LOCKING CONNECTOR, A DESIGN ENHANCEMENT HAS BEEN IMPLEMENTED (B)(4). IN ADDITION, (B)(4) ADJUSTABLE GASTRIC BAND AND REALIZE ADJUSTABLE GASTRIC BANDING PRODUCTS WITH THE PRE-ENHANCEMENT DESIGN WAS A VOLUNTARY RECALLED. A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED AND NO DISCREPANCIES WERE RECORDED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT POST IMPLANT, A LAPAROSCOPIC ADJUSTABLE BAND, THE PATIENT WAS NOT LOSING WEIGHT. FURTHER EVALUATION WAS DONE AND NO LEAK WAS FOUND. THE STRAIN RELIEF WAS FOUND TO HAVE MIGRATED DOWN FROM THE CONNECTION HOUSING ON THE BAND TUBING. ON (B)(6) 2010, THE PORT WAS REPLACED. THE PATIENT HAD 3-4 FILLS. THERE WAS NO ADVERSE PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REALIZE ADJ GASTRIC BAND-C IMPLANT, INTRAGASTRIC FOR MORBID OBESITY LTI OBTECH MEDICAL SARL_ UNK NI

Patients

Seq Age Sex Outcome Treatment
1