FDA Adverse Event Injury Summary report: N

ENDOPATH ETS45

MDR report key: 1890053 · Received November 2, 2010

Report

Report Number
3005075853-2010-06222
Event Type
Injury
Date Received
November 2, 2010
Date of Event
October 4, 2010
Report Date
October 5, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DATE SENT: 11/02/2010. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE ETS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE (USING THE OMEGA LOOP TECHNIQUE), THE SURGEON OBSERVED THE STAPLE LINES WERE NICE AND DRY MAKING THE POUCH AND ALSO AFTER DIVIDING THE INTESTINE. AFTER DIVISION OF THE JEJUNUM, THE SURGEON RETURNED TO THE GJ ANASTOMOSIS FOR THE LEAK TEST. THE SURGEON FOUND THAT THE STAPLE LINES HAD STARTED TO BLEED AND HE NEEDED TO PLACE SEVERAL SUTURES FOR HEMOSTASIS. HE ALSO INSPECTED THE TRANSECTION OF THE JEJUNUM, AND THAT STAPLE LINE HAD ALSO STARTED TO BLEED THROUGH THE STAPLES AND HE USED CLIPS TO GET HEMOSTASIS. THE CONDITION OF THE PT IMMEDIATELY FOLLOWING THE EVENT WAS STABLE. THE CARTRIDGE FOR TRANS OF THE JJ IS WHITE - STILL IN THE JAWS OF THE DEVICE. PT DID NOT EXPERIENCE HOSPITALIZATION LONGER THAN EXPECTED; HOWEVER, PT WAS IN THE OPERATING ROOM AFTER THE PLANNED SURGERY FOR A F/U EXAM VIA SCOPE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS45 GDW STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC NA G4TF21

Patients

Seq Age Sex Outcome Treatment
1 32 YR Other TR45W, LOT # G4T826| 2ND RELOAD PRODUCT # 6R45B