ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2010-21250
- Event Type
- Malfunction
- Date Received
- November 3, 2010
- Date of Event
- May 26, 2009
- Report Date
- August 29, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SUBSEQUENTLY, IT WAS NOTED THAT SHOCK IMPEDANCE MEASUREMENTS HAVE BEEN TRENDING DOWN. THIS WAS OBSERVED DURING A DEVICE CHECK PRIOR TO RADIATION THERAPY, WHERE SHOCK IMPEDANCE MEASUREMENTS WERE IN THE 85 OHM RANGE. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED AS A RESULT OF THIS OBSERVATION. AVAILABLE INFORMATION SUGGESTS THAT THESE PRODUCTS REMAINS IN SERVICE, WITH THE SHOCK IMPEDANCE CONTINUING TO BE MONITORED. THIS EVENT WILL BE UPDATED IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
ADDITIONAL INFORMATION WAS RECEIVED THAT AN INCREASED OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT WAS OBSERVED ONCE AGAIN ON THIS DEVICE SYSTEM. THE LOCAL BOSTON SCIENTIFIC SALES REPRESENTATIVE REPORTED THAT THE PATIENT IMPLANTED WITH THIS DEVICE SYSTEM RECENTLY ENDURED A NON-DEVICE RELATED PROCEDURE AND WAS RECOVERING. FOLLOWING RECOVERY, THE PATIENT'S CARDIOLOGIST PLANS TO BRING THE PATIENT IN FOR A NON-INVASIVE PROGRAMMED STIMULATION (NIPS) PROCEDURE TO TEST THE HIGH VOLTAGE SHOCK INTEGRITY OF THE DEVICE SYSTEM. TO DATE, NO ADVERSE PATIENT EFFECTS WERE REPORTED AS A RESULT OF THIS CLINICAL OBSERVATION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND TRANSVENOUS RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD EXHIBITED A GRADUAL INCREASE IN SHOCK IMPEDANCE MEASUREMENTS FROM 77 OHMS AT IMPLANT, TO ONE MEASUREMENT OF GREATER THAN 125 OHMS. NO NOISE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | T177| 0181 |