FDA Adverse Event Injury Summary report: N

FREESTYLE LIBRE 2

MDR report key: 18899000 · Received March 13, 2024

Report

Report Number
2954323-2024-08490
Event Type
Injury
Date Received
March 13, 2024
Date of Event
February 25, 2024
Report Date
April 19, 2024
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
QLG
PMA / PMN Number
K193371
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. THE DEVICE MANUFACTURING DATE IS UNKNOWN. THE DATE ENTERED IN SECTION H4 IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. THE MOST PROBABLE ROOT CAUSES ASSOCIATED WITH THIS FAILURE MODE ARE DISCONNECTED, FAULTY OR DAMAGED COMPONENTS, SOFTWARE/DATA CORRUPTION, OR MISUSE. HOWEVER, MITIGATIONS ARE IN PLACE TO REDUCE AND PREVENT SUCH ISSUES. ALL VITAL MANUFACTURING STEPS ARE VALIDATED, MONITORED, AND VERIFIED DURING MANUFACTURING TO ENSURE THE SYSTEM IS IN CONFORMANCE WITH THE VERIFIED DESIGN. ALL VITAL FUNCTIONS ARE MONITORED BY THE SYSTEM AND, WHEN NECESSARY, FUNCTION IS SUSPENDED TO SAFEGUARD AGAINST INACCURATE RESULTS. LABELING IS PROVIDED TO INSTRUCT THE USER ON THE INTENDED USE OF ALL VITAL PARTS OF THE SYSTEM TO MINIMIZE MISUSE. THE MOST PROBABLE ROOT CAUSES ASSOCIATED WITH THIS FAILURE MODE ARE SOFTWARE/DATA CORRUPTION, OR MISUSE. HOWEVER, MITIGATIONS ARE IN PLACE TO REDUCE AND PREVENT SUCH ISSUES. ALL VITAL FUNCTIONS ARE MONITORED BY THE SYSTEM AND, WHEN NECESSARY, FUNCTION IS SUSPENDED TO SAFEGUARD AGAINST INACCURATE RESULTS. LABELING IS PROVIDED TO INSTRUCT THE USER ON THE INTENDED USE OF ALL VITAL PARTS OF THE SYSTEM TO MINIMIZE MISUSE. ALL COMPLAINTS AND COMPLAINT TRENDS ARE INVESTIGATED TO DETERMINE IF THERE IS A PRODUCT DEFECT/ DEFICIENCY. IF A PRODUCT DEFECT/ DEFICIENCY IS IDENTIFIED, A RISK EVALUATION IS COMPLETED AND COMPARED TO THE RISK MANAGEMENT REPORT, TO ENSURE THE RISK PROFILE HAS NOT CHANGED. ADDITIONALLY, AS A PART OF ABBOTT¿S POST-MARKET SURVEILLANCE PROCESS, ALL RISK EVALUATIONS WITH ASSOCIATED COMPLAINT DATA ARE REVIEWED ANNUALLY TO DETERMINE IF THE RISK PROFILES HAVE CHANGED AS COMPARED TO THE PRODUCT RISK MANAGEMENT REPORTS. THESE MONITORING PROCESSES ENSURE THAT ALL PRODUCT RISK PROFILES REMAIN ACCEPTABLE AND HAVE A POSITIVE BENEFIT/ RISK RATIO. AT THIS TIME PRODUCT HAS NOT YET BEEN RETURNED FOR THIS COMPLAINT. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND IT HAS BEEN DETERMINED THAT THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. THE DHRS (DEVICE HISTORY REVIEW) FOR THE LIBRE SENSOR AND LIBRE SENSOR KITS WERE REVIEWED AND THE DHRS SHOWED THE LIBRE SENSOR AND LIBRE SENSOR KITS PASSED ALL TESTS PRIOR TO RELEASE. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND FREESTYLE LIBRELINK. CUSTOMER COMPLAINED FOR SIGNAL LOSS. THE REPORTED ISSUE WAS INVESTIGATED AND ATTEMPTED TO REPLICATE USING SIMILAR CONFIGURATION. THE REPORTED ISSUE WAS UNABLE TO BE REPLICATED AND THE SYSTEM FUNCTIONED AS INTENDED. CUSTOMER COMPLAINED REGARDING SIGNAL LOSS MAY OCCURRED FOR MANY REASONS. THERE WERE NO ISSUES IDENTIFIED WITH THE FREESTYLE LIBRELINK SPECIAL EDITION APP DURING REPLICATION THAT WOULD HAVE LED TO THE REPORTED ISSUE AS THE CASE. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

AN ALARM ISSUE WAS REPORTED WITH THE ADC DEVICE IN USE WITH NUU PHONE WITH ANDROID OPERATING SYSTEM VERSION 9. THE CUSTOMER EXPERIENCED A SIGNAL LOSS AND WAS UNABLE TO OBTAIN READINGS AND RECEIVE GLUCOSE ALARMS. AS A RESULT, THE CUSTOMER EXPERIENCED A LOSS OF CONSCIOUSNESS. THE CUSTOMER WAS SEEN AT A HOSPITAL AND A HEALTHCARE PROFESSIONAL (HCP) ADMINISTERED 2 GLUCAGON INJECTIONS FOR THE DIAGNOSIS OF HYPOGLYCEMIA. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

AN ALARM ISSUE WAS REPORTED WITH THE ADC DEVICE IN USE WITH NUU PHONE WITH ANDROID OPERATING SYSTEM VERSION 9 AND 2.10.0.6714 APP VERSION. THE CUSTOMER EXPERIENCED A SIGNAL LOSS AND WAS UNABLE TO OBTAIN READINGS AND RECEIVE GLUCOSE ALARMS. AS A RESULT, THE CUSTOMER EXPERIENCED A LOSS OF CONSCIOUSNESS. THE CUSTOMER WAS SEEN AT A HOSPITAL AND A HEALTHCARE PROFESSIONAL (HCP) ADMINISTERED 2 GLUCAGON INJECTIONS FOR THE DIAGNOSIS OF HYPOGLYCEMIA. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1213032 FREESTYLE LIBRE 2 FLASH GLUCOSE MONITORING SYSTEM QLG ABBOTT DIABETES CARE INC 71992-01

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention