SONARMED
Report
- Report Number
- 1282497-2024-00025
- Event Type
- Malfunction
- Date Received
- March 11, 2024
- Date of Event
- November 7, 2022
- Report Date
- May 8, 2024
- Manufacturer
- SONARMED, INC.
- Product Code
- OQU
- UDI-DI
- 00851334007193
- PMA / PMN Number
- K193058
- Removal / Correction Number
- Z-1520-2024
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SECTION E. MEDTRONIC PERSONNEL REPORTED THE EVENT ON BEHALF OF THE CUSTOMER. H3: A PHOTO OF THE DEVICE HAS BEEN RECEIVED. THE INVESTIGATION IS CURRENTLY IN PROGRESS. THERE IS AN EXISTING PREVIOUS INTERNAL INVESTIGATION REGARDING THE REPORTED ISSUE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
SECTION H9: 806 NUMBER 3009023584-03-08-2024-01-R SECTION H6 EVALUATION CODES UPDATED DUE TO COMPLETION OF INVESTIGATION METHOD - ADD ADDITIONAL CODE RESULT - ADD ADDITIONAL CODE CONCLUSION - REMOVE D15 AND ADD D03 COMPONENT - REMOVE G07003 AND ADD G03012 H3 DEVICE EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED ON ALL INFORMATION RECEIVED. IT WAS REPORTED THAT, WHILE IN USE ON A PREMATURE INFANT, THE SONARMED SENSOR MODEL 3.0 HAD ¿PARTIAL OBSTRUCTION AND DID NOT ALLOW SUCTION CATHETER TO PASS EASILY¿. HE SENSOR WAS REPLACED WITH A NEW SENSOR TO SOLVE THE ISSUE. ONE PHOTO IMAGE WAS PROVIDED FOR ANALYSIS. A VISUAL OBSERVATION OF THE ATTACHED PHOTO APPEARS TO SHOW AN OBSTRUCTION IN THE INNER LUMEN OF THE SENSOR. THE CAUSE OF THE ISSUE WAS ISOLATED TO AN OBSTRUCTION IN THE SENSOR. AN EXISTING CORRECTIVE ACTION ADDRESSES THE ISSUE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED THAT, WHILE IN USE ON A PREMATURE INFANT, THE SONARMED SENSOR MODEL 3.0 HAD ¿PARTIAL OBSTRUCTION AND DID NOT ALLOW SUCTION CATHETER TO PASS EASILY¿. THE SENSOR WAS REPLACED WITH A NEW SENSOR, NO HARM TO PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1370451 | SONARMED | AIRWAY MONITORING SYSTEM | OQU | SONARMED, INC. | AW-S030 | 00851334007193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |