FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET

MDR report key: 18882807 · Received March 11, 2024

Report

Report Number
1024879-2024-00216
Event Type
Malfunction
Date Received
March 11, 2024
Date of Event
February 8, 2024
Report Date
March 20, 2024
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FPA
UDI-DI
50382903673426
PMA / PMN Number
K030573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY MATERIAL #: 367342 LOT/BATCH #: 3324486. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, THIRTY (30) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH SUBJECTED TO DRAW AND RETRACTION LOCKOUT TESTING, NO ISSUES WERE OBSERVED RELATING TO LEAKAGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.

Additional Manufacturer Narrative · 0

D2A. COMMON DEVICE NAME: BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET D2B. MEDICAL DEVICE TYPE: JKA THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.5. PMA / 510(K)#: K220212 H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, ONE (1) DEVICE LEAKED FROM THE TUBING. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, ONE (1) DEVICE LEAKED FROM THE TUBING. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
909509 BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET FPA BECTON, DICKINSON & CO., (BD) 3324486 50382903673426

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown