AQUABEAM ROBOTIC SYSTEM
Report
- Report Number
- 3012977056-2024-00068
- Event Type
- Malfunction
- Date Received
- March 11, 2024
- Date of Event
- February 21, 2024
- Report Date
- April 1, 2024
- Manufacturer
- PROCEPT BIOROBOTICS CORPORATION
- Product Code
- PZP
- PMA / PMN Number
- DEN170024
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ROOT CAUSE OF REPORTED EVENT HAS NOT YET BEEN ESTABLISHED. INVESTIGATION BY MANUFACTURER IS CURRENTLY IN-PROCESS. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
THE AQUABEAM HANDPIECE WAS RETURNED FOR INVESTIGATION. INVESTIGATION INTO THIS ISSUE HAS CONFIRMED THE OCCURRENCE OF "E22 - MOTORPACK" ERRORS. THE ISSUE IS BEING ADDRESSED THROUGH PROCEPT'S QUALITY SYSTEM. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) AB2000-B / SERIAL NUMBER (B)(6) AND AQUABEAM HANDPIECE / LOT NUMBER 23C10968 WAS CONDUCTED, WHICH CONFIRMED THAT THERE WAS ONE (1) NON-CONFORMANCE ISSUED TO THIS LOT DURING THE MANUFACTURING PROCESS THAT COULD POTENTIALLY BE RELATED TO THE REPORTED EVENT. THE AFFECTED UNITS WITHIN THE LOT WERE SEGREGATED AND REWORKED TO ADDRESS THE NON-CONFORMANCE. THE HANDPIECE PASSED FINAL INSPECTION PRIOR TO RELEASE FOR DISTRIBUTION. THE CURRENT USER MANUAL SJ-UM0101-00 REV. B, UM, AQUABEAM ROBOTIC SYSTEM USER MANUAL, US, BAYTECH WAS REVIEWED AND STATES: TABLE 5: SYSTEM DETECTED ERRORS AND FAULTS E22 - MOTORPACK ERROR. RELEASE FOOT PEDAL AND CLICK X. 1) IF ERROR PERSISTS, RECONNECT HANDPIECE TO MOTORPACK. 2) IF ERROR CONTINUES, REPLACE HANDPIECE. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION BECAME AWARE THAT DURING AQUABLTION PROCEDURE, THE AQUABEAM ROBOTIC SYSTEM GENERATED AN "E22 - MOTORPACK ERROR" AND COULD NOT BE CLEARED DESPITE MULTIPLE TROUBLESHOOTING STEPS. AS A RESULT A SECOND AQUABEAM HANDPIECE WAS OPENED AND PROCEDURE WAS COMPLETED SUCCESSFULLY. THE REPORTED EVENT CAUSED A SURGICAL DELAY OF OVER 20 MINUTES. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2454807 | AQUABEAM ROBOTIC SYSTEM | FLUID JET REMOVAL SYSTEM | PZP | PROCEPT BIOROBOTICS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male |