FDA Adverse Event Malfunction Summary report: N

ATHLETIS PTA BALLOON DILATATION CATHETER

MDR report key: 18877813 · Received March 11, 2024

Report

Report Number
2124215-2024-11179
Event Type
Malfunction
Date Received
March 11, 2024
Date of Event
February 19, 2024
Report Date
April 15, 2024
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LIT
UDI-DI
08714729974260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D2B. PRO CODE (PRODUCT CODE)-LIT, DQY DEVICE EVALUATED BY MFR.: THE DEVICE WAS RETURNED FOR ANALYSIS. A VISUAL EXAMINATION IDENTIFIED THAT THE BALLOON WAS NOT FOLDED WHICH INDICATES THAT THE BALLOON WAS SUBJECTED TO POSITIVE PRESSURE. THE RATED BURST PRESSURE FOR THIS DEVICE IS 20 ATMOSPHERES AS PER ATHLETIS SPECIFICATION. THE RETURNED DEVICE WAS ATTACHED TO AN ENCORE INFLATION UNIT. POSITIVE PRESSURE WAS APPLIED WHEN LIQUID WAS OBSERVED TO BE LEAKING FROM A BALLOON PINHOLE LOCATED APPROXIMATELY 13MM DISTAL FROM THE PROXIMAL MARKERBAND. A VISUAL AND TACTILE EXAMINATION IDENTIFIED NO KINKS OR DAMAGE TO THE SHAFT. A VISUAL AND TACTILE EXAMINATION IDENTIFIED NO DAMAGE TO THE TIP.

Additional Manufacturer Narrative · 0

D2B. PRO CODE (PRODUCT CODE)-LIT, DQY.

Description of Event or Problem · 0

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON BURST JUST BEFORE REACHING PRESSURE. THE SAME PRODUCT WAS THEN USED FOR RETREATMENT WITHOUT ANY PROBLEM. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON BURST JUST BEFORE REACHING PRESSURE. THE SAME PRODUCT WAS THEN USED FOR RETREATMENT WITHOUT ANY PROBLEM. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2076126 ATHLETIS PTA BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC CORPORATION 39347-120470 0029389743 08714729974260

Patients

Seq Age Sex Outcome Treatment
1 70 YR Male