FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 18872729 · Received March 11, 2024

Report

Report Number
9610877-2024-51940
Event Type
Malfunction
Date Received
March 11, 2024
Date of Event
February 27, 2024
Report Date
March 11, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EB15-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K200678. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE DISTAL END WITH CCD MODULE FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE DISTAL END WITH CCD MODULE. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE LIGHT GUIDE CABLE FLUID DAMAGE, THE LG CONNECTOR FLUID DAMAGE, THE SHIELD COVER FLUID DAMAGE, THE BENDING RUBBER PERFORATED, THE ANGLE LEVER TRIM CAP MISSING, THE INSERTION FLEXIBLE TUBE CRUSHED, AND THE REMOTE CONTROL BUTTONS WORN OUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
178617 PENTAX VIDEO BRONCHOSCOPE EOQ HOYA CORPORATION PENTAX TOKYO OFFICE EB11-J10

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown