FDA Adverse Event Malfunction Summary report: N

CARDIOMEMS PATIENT ELECTRONIC SYSTEM

MDR report key: 18855345 · Received March 7, 2024

Report

Report Number
3004936110-2024-00351
Event Type
Malfunction
Date Received
March 7, 2024
Date of Event
March 5, 2024
Report Date
April 8, 2024
Manufacturer
ST. JUDE MEDICAL, INC.
Product Code
MOM
UDI-DI
05414734509800
PMA / PMN Number
P100045
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.

Additional Manufacturer Narrative · 0

ONE CARDIOMEMS PATIENT ELECTRONIC SYSTEM WAS RECEIVED FOR EVALUATION.. EXTERNAL AND INTERNAL VISUAL INSPECTIONS OF UNIT REVEALED EXPOSED AND DETACHED WIRING FROM THE HANDHELD POWER CABLE. DURING THE INVESTIGATION, VISUAL INSPECTION REVEALED THAT THE POWER EXTENSION CONNECTOR WAS SEATED IN THE CONNECTOR COVER. IN ADDITION, NO POWER CLIP WAS ATTACHED AND/OR RETURNED. ALL STANDARD ACCESSORIES WERE USED INCLUDING HANDHELD CASING, LCD SCREEN, HANDHELD PCB BOARD, HANDHELD POWER CABLE, AND GOLDEN HANDHELD MEMBRANE BUTTON. NO SOFTWARE MALFUNCTIONS, ERRORS, WARNINGS, OR ANOMALIES WERE OBSERVED DURING UNIT BOOT OR DURING HANDHELD TOUCH SCREEN COMMANDS. THE FIELD REPORTED EVENT OF HANDHELD DEVICE CORD BEING SEVERED AND DETACHED FROM THE PATIENT ELECTRONIC UNIT WAS CONFIRMED. A REVIEW OF THE DHR WAS PERFORMED FOR BATCH #9043533 AND RULED OUT THE POSSIBILITY OF A MANUFACTURING RELATED ISSUE CAUSING OR CONTRIBUTING TO THE REPORTED COMPLAINT ISSUE, AS NO NON-CONFORMANCE RELATED TO THE REPORTED ISSUE WAS FOUND IN THE BATCH RECORDS REVIEWED.

Description of Event or Problem · 0

THE PATIENT REPORTED EXPOSED WIRES ON THE HANDHELD DEVICE CABLE. THE PATIENT ELECTRONIC UNIT WAS REPLACED AND THERE WERE NO ADVERSE CONSEQUENCES REPORTED BY THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
908326 CARDIOMEMS PATIENT ELECTRONIC SYSTEM SYSTEM, HEMODYNAMIC, IMPLANTABLE MOM ST. JUDE MEDICAL, INC. CM1100 9043533 05414734509800

Patients

Seq Age Sex Outcome Treatment
1 43 YR Male