CE INFUSOR, LV 5 ML/H,12 PACK
Report
- Report Number
- 6000001-2010-04581
- Event Type
- Malfunction
- Date Received
- October 28, 2010
- Date of Event
- August 7, 2010
- Report Date
- August 11, 2010
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL EXAMINATION WAS PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A RUPTURED RESERVOIR. THE DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH IDC-CAPA-(B)(4). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.
IT WAS REPORTED TO BAXTER (B)(4) THAT THE RESERVOIR OF AN INFUSOR LV 5 DEVICE RUPTURED DURING FILLING. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CE INFUSOR, LV 5 ML/H,12 PACK | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 09D017 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |