FDA Adverse Event Malfunction Summary report: N

CE INFUSOR, LV 5 ML/H,12 PACK

MDR report key: 1885345 · Received October 28, 2010

Report

Report Number
6000001-2010-04580
Event Type
Malfunction
Date Received
October 28, 2010
Date of Event
August 8, 2010
Report Date
August 11, 2010
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. A VISUAL EXAMINATION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF BACKFLOW FROM THE FILLING PORT. THE ROOT CAUSE COULD WAS DETERMINED TO BE A PARTICLE TRAPPED BETWEEN THE CHECKBAND AND THE STRESSMEMBER. THIS DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT AN INFUSOR LV 5 DEVICE WAS LEAKING FROM THE FILL PORT DURING FILLING. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CE INFUSOR, LV 5 ML/H,12 PACK PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 09C010

Patients

Seq Age Sex Outcome Treatment
1