Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE WAS REPLACED AND THE SOFTWARE RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO SAVE IMAGES. NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPY XRAY JAA GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE) 8800

Patients

Seq Age Sex Outcome Treatment
1