FDA Adverse Event Malfunction Summary report: N

CARESTREAM DRX-EVOLUTION SYSTEM

MDR report key: 18852528 · Received March 6, 2024

Report

Report Number
1317307-2024-00002
Event Type
Malfunction
Date Received
March 6, 2024
Date of Event
February 16, 2024
Report Date
March 4, 2024
Manufacturer
CARESTREAM HEALTH INC.
Product Code
KPR
UDI-DI
60889971739193
PMA / PMN Number
K091889
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CARESTREAM HEALTH (CSH) HAS INVESTIGATED THIS ISSUE. THE INVESTIGATION DETERMINED THAT THE SYSTEM MIGHT APPLY THE PREVIOUS TECHNIQUES DATA TO THE CURRENT VIEW FOR USE, RESULTING IN A HIGHER EXPOSURE. CARESTREAM IS WORKING THROUGH PROVIDING A SOFTWARE SOLUTION. A WORK-AROUND HAS BEEN PROVIDED TO THE CUSTOMER UNTIL THE NEXT RELEASE OF SOFTWARE WHICH CONTAINS A SOLUTION TO ADDRESS THIS TYPE OF ISSUE. USERS HAVE VISIBILITY TO TECHNIQUES ON THE USER INTERFACE; IT IS EXPECTED THAT A SKILLED AND TRAINED RADIOLOGICAL HEALTH CARE PROFESSIONAL VERIFIES ALL TECHNIQUES PRIOR TO EVERY EXPOSURE. PER RADIATION SAFETY, BASED ON THE TECHNIQUES USED, THE RISK TO THIS PATIENT FROM THESE IMAGES IS MINIMAL. THE ADDITIONAL RADIATION DOSE IS WELL BELOW THAT WHICH WOULD RESULT IN ANY PROMPT EFFECTS LIKE SKIN ERYTHEMA, AND THE ADDITIONAL RISK FROM STOCHASTIC EFFECTS IS NEGLIGIBLE; HOWEVER, CSH WILL REPORT THIS INCIDENT TO ERR ON THE SIDE OF CAUTION. IF THE ISSUE WERE TO RE-OCCUR, IT MAY REQUIRE A REPEAT IMAGE/RE-EXPOSURE. A RE-EXPOSURE WILL NOT LEAD TO A SERIOUS INJURY. THERE ARE NO FURTHER ACTIONS TO BE TAKEN BY CARESTREAM HEALTH RELATED TO THIS ISSUE. CARESTREAM HAS COMPLETED THIS INVESTIGATION.

Description of Event or Problem · 0

ON (B)(6) 2024, CARESTREAM HEALTH (CSH) WAS INFORMED OF AN INCIDENT RELATED TO THE DRX-EVOLUTION SYSTEM THAT OCCURRED ON (B)(6) 2024. THE PROBLEM DESCRIPTION OUTLINES A RADIATION DOSE HIGHER THAN EXPECTED. NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175435 CARESTREAM DRX-EVOLUTION SYSTEM CARESTREAM DRX-EVOLUTION SYSTEM KPR CARESTREAM HEALTH INC. 60889971739193

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other