FDA Adverse Event Injury Summary report: N

INTERTAN

MDR report key: 1884924 · Received October 28, 2010

Report

Report Number
1020279-2010-00306
Event Type
Injury
Date Received
October 28, 2010
Date of Event
August 25, 2010
Report Date
March 10, 2011
Manufacturer
SMITH-NEPHEW INC.
Product Code
LYT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FROM THE ANALYSES CONDUCTED DURING THIS INVESTIGATION, IT WAS SHOWN THAT THE T-HANDLE FRACTURED IN THE 'EAR' OF THE HANDLE IN THE CLAMPING JAW, LIKELY FROM STATIC BENDING OR TORSIONAL OVERLOAD. AN OVERLOAD FRACTURE CAN OCCUR IF THE MECHANICAL LOADS APPLIED TO THE DEVICE EXCEED THE STRENGTH OF THE MATERIAL. NO MATERIAL OR MANUFACTURING DEVIATIONS WERE FOUND DURING THIS INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGERY TIME WAS EXTENDED DUE TO A BROKEN T-HANDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERTAN T-HANDLE LYT SMITH-NEPHEW INC. 09ACT0010

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization