FDA Adverse Event
Injury
Summary report: N
INTERTAN
MDR report key: 1884924
·
Received October 28, 2010
Report
- Report Number
- 1020279-2010-00306
- Event Type
- Injury
- Date Received
- October 28, 2010
- Date of Event
- August 25, 2010
- Report Date
- March 10, 2011
- Manufacturer
- SMITH-NEPHEW INC.
- Product Code
- LYT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FROM THE ANALYSES CONDUCTED DURING THIS INVESTIGATION, IT WAS SHOWN THAT THE T-HANDLE FRACTURED IN THE 'EAR' OF THE HANDLE IN THE CLAMPING JAW, LIKELY FROM STATIC BENDING OR TORSIONAL OVERLOAD. AN OVERLOAD FRACTURE CAN OCCUR IF THE MECHANICAL LOADS APPLIED TO THE DEVICE EXCEED THE STRENGTH OF THE MATERIAL. NO MATERIAL OR MANUFACTURING DEVIATIONS WERE FOUND DURING THIS INVESTIGATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT SURGERY TIME WAS EXTENDED DUE TO A BROKEN T-HANDLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTERTAN | T-HANDLE | LYT | SMITH-NEPHEW INC. | 09ACT0010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |