INOMAX DS (DELIVERY SYSTEM)
Report
- Report Number
- 3004531588-2010-00097
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Date of Event
- July 20, 2010
- Report Date
- October 15, 2010
- Manufacturer
- INO THERAPEUTICS LLC/IKARIA
- Product Code
- MRN
- PMA / PMN Number
- K061901
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
ON (B)(6) 2010, A RESPIRATORY THERAPIST REPORTED FLUCTUATING READINGS ON THE INOMAX DS (B)(4). NO IMPACT TO PT. EVAL SUMMARY: THE ROOT CAUSE OF THE DELIVERY FAILURE INVOLVES FRETTING CORROSION OF CONDUCTING TRACES IN AN INTERNAL RIBBON CABLE. THE FRETTING CORROSION CAN CAUSE INTERMITTENT HIGH RESISTANCE CORRECTION TO THE CABLE'S CONNECTOR, CAUSING FLUCTUATING MONITORED (NO) READINGS. THE INTERNAL RIBBON CABLE WAS REPLACED AND IT WAS CONFIRMED THE MONITORED FLUCTUATING MONITORED (NO) READINGS STOPPED AND WERE CORRECTED. IT IS IMPORTANT TO NOTE THAT THE MONITORED (NO) LEVEL WOULD BE FLUCTUATING IN THIS CASE AND NOT THE ACTUAL (NO) DELIVERED.
ON (B)(6) 2010, A RESPIRATORY THERAPIST REPORTED FLUCTUATING READINGS ON THE INOMAX DS (B)(4). THE RESPIRATORY THERAPIST STATES THERE WAS NO HARM TO PT AND NO ADVERSE EVENT OCCURRED. THE DEVICE WAS REPLACED WITH ANOTHER UNIT. THE DEVICE WAS REMOVED FROM SERVICE BY THE CUSTOMER AND RETURNED TO THE COMPANY FOR INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INOMAX DS (DELIVERY SYSTEM) | APPARATUS, NITRIC OXIDE DELIVERY | MRN | INO THERAPEUTICS LLC/IKARIA | 10003 |