FDA Adverse Event Injury Summary report: N

KIT IMPLANTABLE SLIM TIP LEAD, 50CM

MDR report key: 18847718 · Received March 6, 2024

Report

Report Number
1627487-2024-07367
Event Type
Injury
Date Received
March 6, 2024
Date of Event
February 6, 2024
Report Date
October 28, 2024
Manufacturer
ABBOTT MEDICAL
Product Code
PMP
UDI-DI
05415067027153
PMA / PMN Number
P150004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT ESTIMATED.

Additional Manufacturer Narrative · 0

DATE OF EVENT ESTIMATED. THE ALLEGATION IS AGAINST [1] OF [2] [LEAD]; HOWEVER, IT IS UNKNOWN WHICH [LEAD(S)], THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: LEAD, MODEL: MN10450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 9042817.

Additional Manufacturer Narrative · 0

DATE OF EVENT ESTIMATED. CORRECTION: SECTION B5 CML VEBIAGE WAS INADVERTENTLY ADDED ON THE INITIAL REPORT. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS.

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED STATING THE PATIENT WAS EXPERIENCING SHOCKING AT THE IPG SITE. SURGICAL INTERVENTION TOOK PLACE WHERE THE IPG AND LEAD WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE. EFFECTIVE STIMULATION WAS RESTORED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DEVICE COULD NOT DELIVER THE DESIRED STRENGTH. UPON FURTHER INVESTIGATION THE SYSTEM DIAGNOSTICS RECORDED HIGH IMPEDANCES ON THE LEAD AND THAT THE LEAD HAD MIGRATED, WHICH WAS CONFIRMED VIA X-RAY. AS A RESULT, THE PATIENT IS EXPERIENCING INEFFECTIVE STIMULATION. SURGICAL INTERVENTION MAY TAKE PLACE AT A FUTURE DATE TO ADDRESS THE ISSUE. INVESTIGATION WAS UNABLE TO DETERMINE WHICH OF THE LEADS ATTRIBUTED TO THE EVENT. THE ALLEGATION IS AGAINST [1] OF [2] [LEAD]; HOWEVER, IT IS UNKNOWN WHICH [LEAD(S)], THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: LEAD, MODEL: MN10450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 9042817.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
118378 KIT IMPLANTABLE SLIM TIP LEAD, 50CM SCS LEAD PMP ABBOTT MEDICAL MN10450-50A 9042817 05415067027153

Patients

Seq Age Sex Outcome Treatment
1 NA Female Other SCS IPG| SCS LEAD