FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFSUION PUMP
MDR report key: 1884695
·
Received October 25, 2010
Report
- Report Number
- 2531779-2010-01883
- Event Type
- Malfunction
- Date Received
- October 25, 2010
- Date of Event
- September 24, 2010
- Report Date
- September 24, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. DURING EVAL, THE 'UP' 'DOWN' AND 'OK' BUTTONS REQUIRED MULTIPLE PRESSES TO RESPOND. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
THE PT'S FATHER REPORTED THAT THE 'UP' 'DOWN' AND 'OK' BUTTONS. NO TRAUMA TO THE PUMP OR EXPOSURE TO MOISTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFSUION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |