FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFSUION PUMP

MDR report key: 1884695 · Received October 25, 2010

Report

Report Number
2531779-2010-01883
Event Type
Malfunction
Date Received
October 25, 2010
Date of Event
September 24, 2010
Report Date
September 24, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. DURING EVAL, THE 'UP' 'DOWN' AND 'OK' BUTTONS REQUIRED MULTIPLE PRESSES TO RESPOND. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PT'S FATHER REPORTED THAT THE 'UP' 'DOWN' AND 'OK' BUTTONS. NO TRAUMA TO THE PUMP OR EXPOSURE TO MOISTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFSUION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 15 YR