FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1884688
·
Received October 25, 2010
Report
- Report Number
- 2531779-2010-01858
- Event Type
- Malfunction
- Date Received
- October 25, 2010
- Date of Event
- September 23, 2010
- Report Date
- September 23, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE PEELING OFF THE PUMP. THE 'UP' AND 'DOWN' BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING. THE UP AND DOWN BUTTON CONTACTS WERE FOUND TO BE MISALIGNED. VISUAL INSPECTION REVEALED A BATTERY COMPARTMENT CRACK.
Description of Event or Problem · 1
THE PT'S FATHER REPORTED THAT THE KEYPAD IS DETACHING FROM THE PUMP NEAR THE DOWN ARROW AND THE DOWN ARROW IS NOT ALWAYS RESPONDING. THE PT REPORTS NOT EXPOSING THE PUMP TO WATER OR CLEANING PRODUCTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |