FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1884688 · Received October 25, 2010

Report

Report Number
2531779-2010-01858
Event Type
Malfunction
Date Received
October 25, 2010
Date of Event
September 23, 2010
Report Date
September 23, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE PEELING OFF THE PUMP. THE 'UP' AND 'DOWN' BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING. THE UP AND DOWN BUTTON CONTACTS WERE FOUND TO BE MISALIGNED. VISUAL INSPECTION REVEALED A BATTERY COMPARTMENT CRACK.

Description of Event or Problem · 1

THE PT'S FATHER REPORTED THAT THE KEYPAD IS DETACHING FROM THE PUMP NEAR THE DOWN ARROW AND THE DOWN ARROW IS NOT ALWAYS RESPONDING. THE PT REPORTS NOT EXPOSING THE PUMP TO WATER OR CLEANING PRODUCTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 14 YR