FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1884686 · Received October 25, 2010

Report

Report Number
2531779-2010-01848
Event Type
Malfunction
Date Received
October 25, 2010
Date of Event
September 23, 2010
Report Date
September 23, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE 'DOWN' BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONDING. THE DOWN BUTTON CONTACT WAS FOUND TO BE MISALIGNED. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PT REPORTED THAT THE 'UP' BUTTON STICKS AND CONTINUES TO SCROLL AFTER RELEASING IT. NO SIGNS OF DAMAGE OR PEELING TO THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR