FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 1884612 · Received October 19, 2010

Report

Report Number
2953200-2010-02002
Event Type
Injury
Date Received
October 19, 2010
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: (GRAFT OCCLUSION), (NARROW AORTIC BIFURCATION), (USE OF THE DEVICE TO TREAT AN ILIAC ANEURYSM WITHOUT AN ABDOMINAL AORTIC ANEURYSM). CONCLUSIONS: (NARROW AORTIC BIFURCATION), (USE OF THE DEVICE TO TREAT AN ILIAC ANEURYSM WITHOUT AN ABDOMINAL AORTIC ANEURYSM; USE OF THE DEVICE IN A NARROW AORTIC BIFURCATION).

Description of Event or Problem · 1

A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 6.7 CM RIGHT ILIAC ARTERY ANEURYSM APPROX 2 MONTHS AGO. THE AORTIC BIFURCATION MEASURED 19 MM IN DIAMETER, AND THE VESSELS HAD MILD TORTUOSITY AND MILD CALCIFICATION. IT WAS REPORTED THAT APPROX 1 MONTH POST-IMPLANTATION, A FOLLOW-UP CT DEMONSTRATED THAT THERE WAS AN OCCLUSION IN THE BIFURCATED STENT GRAFT IN THE UNSUPPORTED AREA. THE PHYSICIAN ELECTED TO INTERVENE BY IMPLANTING AN ILIAC LIMB AND A BALLOON-EXPANDABLE BARE METAL STENT. HOWEVER, THE IMPLANTATION OF THE ILIAC LIMB AND BALLOON-EXPANDABLE STENT ENDED UP OCCLUDING THE CONTRALATERAL ILIAC STENT GRAFT (MFR REPORT #2953200-2010-02003). THE PHYSICIAN THEN ELECTED TO PERFORM A FEMORAL-TO-FEMORAL BYPASS, WHICH SUCCESSFULLY RESOLVED THE OCCLUSION. THE PHYSICIAN STATED THAT THE CAUSE OF THE OCCLUSION IS MOST LIKELY DUE TO THE SMALL IN DIAMETER AORTIC BIFURCATION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00447718

Patients

Seq Age Sex Outcome Treatment
1 82 YR Required Intervention