FDA Adverse Event Malfunction Summary report: N

HAMILTON MEDICAL AG

MDR report key: 18845342 · Received March 6, 2024

Report

Report Number
3001421318-2023-40870
Event Type
Malfunction
Date Received
March 6, 2024
Date of Event
January 26, 2023
Report Date
March 5, 2024
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
00730002856789
PMA / PMN Number
K181216
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT. THE ROOT CAUSE OF THE VENTILATOR WAS A MALFUNCTION OF THE CHECK VALVE ASSEMBLY. THERE WAS NO PATIENT OR USER HARM. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT.

Description of Event or Problem · 0

RELEASE VALVE DEFECTIVE WAS SHOWING ON THE DISPLAY, REPLACEMENT OF THE CHECK VALVE ASSEMBLY RESOLVED THE ISSUE. SERVICE SOFTWARE AND FUNCTION CHECKS ALL PASS. FROM: 161192: 04: 2034 TO MSP161192: 04: 2031.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
119623 HAMILTON MEDICAL AG HAMILTON-T1 CBK HAMILTON MEDICAL AG HAMILTON-T1 00730002856789

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown