HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2024-00519
- Event Type
- Malfunction
- Date Received
- March 6, 2024
- Date of Event
- February 21, 2024
- Report Date
- January 5, 2025
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 07630002800747
- PMA / PMN Number
- K181216
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
HAMILTON MEDICAL AG CASE NUMBER IS: CER (B)(4). FOLLOW-UP 1 - CORRECTED INFORMATION: **UDI RELATED DATA QUALITY UPDATES ONLY** FIELD D4 WAS UPDATED WITH FULL UDI INFORMATION.
HAMILTON MEDICAL AG CASE NUMBER IS: (B)(4).
HAMILTON MEDICAL AG CASE NUMBER IS: (B)(4). A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS PREPARED FOR TREATMENT. BASED ON THE INFORMATION FOUND, THE ROOT CAUSE AND THE CORRECTION COULD NOT BE DETERMINED. THE CUSTOMER DID NOT PROVIDE SUFFICIENT INFORMATION. THERE WAS NO PATIENT OR USER HARM REPORTED.
THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: DURING START UP THE TF 232056 (PLAUSIBILITY BETWEEN PAMBIENT AND PFILTER FAILED) OCCURS AND SELFTEST FAILS. NO HEALTH CONSEQUENCES OR IMPACT.
THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: DURING START UP THE TF 232056 (PLAUSIBILITY BETWEEN PAMBIENT AND PFILTER FAILED) OCCURS AND SELFTEST FAILS. NO HEALTH CONSEQUENCES OR IMPACT.
THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: DURING START UP THE TF 232056 (PLAUSIBILITY BETWEEN PAMBIENT AND PFILTER FAILED) OCCURS AND SELFTEST FAILS. NO HEALTH CONSEQUENCES OR IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 119383 | HAMILTON MEDICAL AG | HAMILTON-C1 | CBK | HAMILTON MEDICAL AG | HAMILTON-C1 | 07630002800747 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |