EXCELSIUS GPS
Report
- Report Number
- 3004142400-2021-00273
- Event Type
- Malfunction
- Date Received
- March 6, 2024
- Date of Event
- October 12, 2021
- Report Date
- March 6, 2024
- Manufacturer
- GLOBUS MEDICAL, INC.
- Product Code
- OLO
- PMA / PMN Number
- K171651
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- 003
Narratives
THIS REPORT IS BEING RESUBMITTED FOR AN INCIDENT THAT OCCURRED EARLIER. INVESTIGATION REVEALED THAT THERE WAS NO SYSTEM MALFUNCTION. ENGINEERING EVALUATIONS REVEALED THAT THERE WAS NO SYSTEM MALFUNCTION. INVESTIGATION OF THE CASE LOGS REVEALED THAT THE ROOT CAUSE WAS USER TECHNIQUE. WHEN LOOKING AT THE PREOPERATIVE MERGE, THERE IS A SHIFT PRESENT IN THE LATERAL MERGE. MERGE SHIFTS CAN CAUSE NAVIGATION INTEGRITY ISSUES AND CAN LEAD TO THE MISPLACEMENT OF SCREWS. THE USER MUST VERIFY THE MERGE BEFORE PROCEEDING TO NAVIGATION. THE TWO SCREWS PLACED USING THE SYSTEM WERE LEFT AS IS. THERE WAS NO ADVERSE EFFECT TO THE PATIENT. THE CAUSE OF THE REPORTED ISSUE WAS TRACED TO USER TECHNIQUE.
WE NEEDED TO ABORT THE CASE (FRACTURE TH7, PRE-OP WORKFLOW) WITH EGPS. AFTER PLANNING WITH CREO MIS, C-ARM SHOTS WERE TAKEN IN AP AND LATERAL. SYSTEM MERGED THE SHOTS WITH PREOP SCAN BUT MERGE WAS OFF. WE REPEATED THE REGISTRATION 2 TIMES, AP AND LATERAL SHOTS, WITH THE SAME RESULTS. THE SHIFT WAS MOSTLY VISIBLE AT TH6. WE PLANNED ONE ADDITIONAL LEVEL ABOVE AND BELOW OUR POINT OF INTEREST. THIS TIME THE MERGE LOOKED GOOD. BUT DURING THE FIRST SCREW PLACEMENT(TH6 L) WE SAW THE LITTLE(SOME MM) INACCURACY IN THE DEPTH. THE SURGEON PUT TWO SCREWS (TH6 L AND R) WITH EGPS AND THE LAST TWO SCREWS IN TRADITIONAL WAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 102226 | EXCELSIUS GPS | SYSTEM, EXCELSIUS GPS, 120V | OLO | GLOBUS MEDICAL, INC. | 999.893 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |