FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1884165 · Received October 27, 2010

Report

Report Number
1423500-2010-04976
Event Type
Malfunction
Date Received
October 27, 2010
Date of Event
September 21, 2010
Report Date
October 5, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL DEVICE WAS RECEIVED AND EVALUATED. THE RITE (RETURN INSTRUMENT TEST/EVALUATION) TEST WAS PERFORMED WHEN THE DEVICE WAS RETURNED TO THE BAXTER (B)(4) FACILITY FOR EVALUATION. THE DEVICE PASSED THE HOMECHOICE RITE FUNCTIONAL TEST ((B)(4)) AND THE HOMECHOICE RITE ELECTRICAL TEST ((B)(4)). DURING THE PAL EVALUATION A SHORT SIMULATED THERAPY WAS PERFORMED AND COMPLETED SUCCESSFULLY. CRT (CYCLER REMOTE TOOLBOX) SOFTWARE INDICATED ALL PRESSURES WERE CORRECT AND STABLE. THE COVER WAS OPENED AND AN INTERNAL INSPECTION PERFORMED. NO PROBLEMS WERE REVEALED DURING THIS INSPECTION AND ALL CONNECTIONS WERE CORRECT AND SECURE. THE IIPV WAS CONFIRMED IN THE LOGS, BUT NOT DUPLICATED DURING PAL EVALUATION. THE ROOT CAUSE IS INSUFFICIENT DRAIN; ONE OR MORE CYCLES ADVANCE TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SERVICE HISTORY REVIEW REVEALED THE PREVIOUS RETURN OF THE DEVICE WAS NOT FOR ANY ISSUES RELATED TO IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE DEVICE, A BAXTER TECHNICIAN IDENTIFIED AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 3. THE DRAIN VOLUME WAS 3418 ML. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 58 YR