COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL
Report
- Report Number
- 6000001-2010-04547
- Event Type
- Malfunction
- Date Received
- October 27, 2010
- Date of Event
- August 2, 2010
- Report Date
- August 25, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A FAILURE CODE 810:11. THE ROOT CAUSE COULD NOT BE DETERMINED COULD NOT BE DETERMINED AT THIS TIME. NO REPAIRS WERE PERFORMED AT THIS TIME, AS THIS IS A BAXTER OWNED DEVICE. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS NOT PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
THE CUSTOMER REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:11. IT IS UNKNOWN WHEN THE REPORTED CONDITION WAS IDENTIFIED. BAXTER'S REVIEW OF THE DEVICE EVENT HISTORY DETERMINED THAT THIS CONDITION INTERRUPTED DELIVERY. THERE WAS NO DOCUMENTED PATIENT INJURY OR MEDICAL INTERVENTION RELATED TO THE REPORTED CONDITION. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.09.90, WHICH IS CLASSIFIED AS REMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.