FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO

MDR report key: 1884031 · Received October 14, 2010

Report

Report Number
2953200-2010-01953
Event Type
Injury
Date Received
October 14, 2010
Date of Event
September 14, 2010
Report Date
September 14, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVAL RESULTS: (ENDOLEAK), (DISEASE PROGRESSION).

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM 113 MONTHS AGO. (MFR REPORT #2953200-2010-01876). THE AORTIC NECK IS 5 CM IN LENGTH. THE ANEURYSM WAS 6.5 CM IN DIAMETER AT TIME OF IMPLANT AND IN THE LEFT ILIAC ARTERY ,THERE WAS A 4.0 CM ANEURYSM. CURRENTLY, THE ABDOMINAL ANEURYSM IS SMALLER IN DIAMETER THAN AT THE TIME OF IMPLANT, IT IS 3.8 CM IN DIAMETER AND THE ILIAC ARTERY ANEURYSM IS GONE. THE VESSEL HAS REMODELED AND RESULTED IN STENT GRAFT MIGRATION. THERE IS ANOTHER ANEURYSM 3.5 CM IN DIAMETER ABOVE THE STENT GRAFT UP TO THE RENAL ARTERIES. THE PHYSICIAN ELECTED TO TREAT THE PATIENT WITH AN ANEURX AORTIC CUFF. THE CUFF WAS IMPLANTED HOWEVER; THERE WAS STILL A PROXIMAL TYPE I ENDOLEAK. THE PHYSICIAN THEN IMPLANTED A TALENT AORTIC CUFF HOWEVER; THE ENDOLEAK STILL WAS NOT RESOLVED. (MFR 2953200-2010-01954). THE PT'S BLOOD PRESSURE DROPPED AND THE PHYSICIAN SUSPECTED A POSSIBLE RUPTURE, HOWEVER, THE PHYSICIAN CONVERTED THE PATIENT TO AN OPEN REPAIR AND FOUND THAT THERE WAS NOT A RUPTURE. THE CAUSE OF THE BLOOD PRESSURE DROPPING IS UNK. THE EXPLANTED STENT GRAFTS WERE DISCARDED BY THE USER FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00312642

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention