FDA Adverse Event Malfunction Summary report: N

TRAUMA STRETCHER

MDR report key: 1884005 · Received October 22, 2010

Report

Report Number
3006697241-2010-00031
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 28, 2010
Report Date
September 28, 2010
Manufacturer
HILL-ROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICAL SUPPORT INSTRUCTED THE ACCOUNT TO ADJUST THE BRAKE LINKAGE TURNBUCKLE. THE ACCOUNT HAS NOT RETURNED HILL-ROM'S ATTEMPTS TO DETERMINE THE RESOLUTION OF THE ALLEGED MALFUNCTION.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE HEAD END CASTERS WILL NOT LOCK. THE CASTERS ARE ROTATING AND SWIVELING WHEN IN BRAKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRAUMA STRETCHER WHEELED STRETCHER FPO HILL-ROM DE MEXICO S DE RL DE CV 8040

Patients

Seq Age Sex Outcome Treatment
1