FDA Adverse Event
Malfunction
Summary report: N
COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL
MDR report key: 1883949
·
Received October 21, 2010
Report
- Report Number
- 2023826-2010-01063
- Event Type
- Malfunction
- Date Received
- October 21, 2010
- Date of Event
- August 20, 2010
- Report Date
- September 22, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P990013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVAL: METHOD (OTHER): WORK ORDER SEARCH. RESULTS (OTHER) - VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THE LENS WAS SPLIT AND A PIECE OF THE HAPTIC/OPTIC WAS TORN OFF AND MISSING. THERE WAS A CLEAR SURGICAL RESIDUE ON THE LENS. A PARENT/CHILD WORK ORDER SEARCH WAS PERFORMED AND THERE WERE NO SIMILAR COMPLAINTS FOUND. CONCLUSION - BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVAL OF THE RETURNED PRODUCT, WE WERE UNABLE TO DETERMINE A SPECIFIC ROOT CAUSE OF THE EVENT. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT A HAPTIC TORE AS THE SURGEON INSERTED A CC4204A COLLAMER PLATE LENS. THE LENS WAS REMOVED AND ANOTHER SAME MODEL LENS WAS IMPLANTED WITHOUT ANY PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | CC4204A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | NANOPOINT INJECTOR, LOT # UNK| NANOPOINT CARTRIDGE, LOT # UNK |