FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1883923
·
Received October 22, 2010
Report
- Report Number
- 1824206-2010-10625
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- September 28, 2010
- Report Date
- September 28, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH CLEANED AND RE-SEATED THE FOOT DOWN PILOT VALVE AND LET THE BED SIT UN-OCCUPIED FOR 3 HRS WITH A 50LBS WEIGHT ON THE FOOT SECTION. THE FOOT SECTION DID NOT DRIFT DOWN AT THAT TIME.
Description of Event or Problem · 1
THE NURSE REPORTED THAT THE FOOT OF THE TOTAL CARE BED GOES DOWN SLOWLY ON ITS OWN OVER A 24 HR PERIOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |