FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1883923 · Received October 22, 2010

Report

Report Number
1824206-2010-10625
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 28, 2010
Report Date
September 28, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH CLEANED AND RE-SEATED THE FOOT DOWN PILOT VALVE AND LET THE BED SIT UN-OCCUPIED FOR 3 HRS WITH A 50LBS WEIGHT ON THE FOOT SECTION. THE FOOT SECTION DID NOT DRIFT DOWN AT THAT TIME.

Description of Event or Problem · 1

THE NURSE REPORTED THAT THE FOOT OF THE TOTAL CARE BED GOES DOWN SLOWLY ON ITS OWN OVER A 24 HR PERIOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1