FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1883870
·
Received October 22, 2010
Report
- Report Number
- 3015876-2010-01174
- Event Type
- Malfunction
- Date Received
- October 22, 2010
- Date of Event
- September 24, 2010
- Report Date
- September 24, 2010
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE TO OPERATE ON AC OR DC POWER. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED POWER SUPPLY ASSEMBLY AT THE FAILURE ANALYSIS CENTER AND VERIFIED THE REPORTED FAILURE. HOWEVER, ADD'L EXTENSIVE TESTING WAS UNABLE TO FURTHER DETERMINE A CONCLUSIVE COMPONENT CAUSE OF THE FAILURE.
Description of Event or Problem · 1
DURING A MORNING CHECK, THE FACILITY NURSE FOUND THAT THE DEVICE WOULD NOT TURN ON AC OR DC POWER. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |