FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1883870 · Received October 22, 2010

Report

Report Number
3015876-2010-01174
Event Type
Malfunction
Date Received
October 22, 2010
Date of Event
September 24, 2010
Report Date
September 24, 2010
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE TO OPERATE ON AC OR DC POWER. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED POWER SUPPLY ASSEMBLY AT THE FAILURE ANALYSIS CENTER AND VERIFIED THE REPORTED FAILURE. HOWEVER, ADD'L EXTENSIVE TESTING WAS UNABLE TO FURTHER DETERMINE A CONCLUSIVE COMPONENT CAUSE OF THE FAILURE.

Description of Event or Problem · 1

DURING A MORNING CHECK, THE FACILITY NURSE FOUND THAT THE DEVICE WOULD NOT TURN ON AC OR DC POWER. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA