BUBBLE DETECTOR CABLE - ADULT
Report
- Report Number
- 2184009-2010-00024
- Event Type
- Injury
- Date Received
- October 18, 2010
- Date of Event
- September 17, 2010
- Report Date
- September 17, 2010
- Manufacturer
- MEDTRONIC PERFUSION SYSTEMS
- Product Code
- DWA
- PMA / PMN Number
- K051303
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). ANALYSIS: EVALUATION OF THE BIOCONSOLE FOUND THAT THIS UNIT PERFORMED AS INTENDED WITH NO ISSUES. THE BUBBLE DETECTOR WAS RETURNED TO MEDTRONIC FOR ANALYSIS. VISUAL INSPECTION SHOWED NO OUTWARD SIGNS OF DAMAGE. FUNCTIONAL TESTING FOUND THAT THE BUBBLE DETECTOR WAS NOT OVERLY SENSITIVE. THE DEVICE OPERATED WITH SPECIFICATION. IN AN EFFORT TO REPRODUCE AN INACCURATE BUBBLE ALARM, THE TUBING WAS MOVED WITHIN THE SENSOR; HOWEVER, THIS REQUIRED A SIGNIFICANT AMOUNT OF MOVEMENT TO TRIGGER. DEVICE HISTORY REVIEW IS PENDING. A SUPPLEMENTAL REPORT WILL BE FILED ONCE COMPLETE. CONCLUSION: BASED ON THE AVAILABLE INFORMATION, IT WAS CONCLUDED THAT THIS EVENT WAS CAUSED BY OPERATIONAL CONTEXT. IT IS POSSIBLE THAT THE BUBBLE ALARM TRIGGERED AN ALARM IF THE TUBING WAS MOVED WITHIN THE SENSOR. DURING TESTING, THE TUBING HAD TO BE MOVED SIGNIFICANTLY TO REPRODUCE THIS INACCURATE BUBBLE ALARM TO BE TRIGGERED.
MEDTRONIC RECEIVED INFORMATION THAT DURING BYPASS, THE BUBBLE DETECTOR ALARM WENT OFF. THE BIOCONSOLE STOPPED FLOW IN RESPONSE TO THE BUBBLE DETECTOR ALARM, AS PROGRAMMED. THE BUBBLE SENSOR WAS REPOSITIONED, AND THE BIOCONSOLE RESTARTED. THE BUBBLE DETECTOR AGAIN WENT OFF. THE SENSOR AND SENSOR LINES WERE EXAMINED AND PROPER POSITIONING WAS VERIFIED. THE BIOCONSOLE WAS AGAIN RESTARTED. THIS OCCURRED 4 TIMES TOTAL. IT WAS REPORTED THAT DURING THE TROUBLESHOOTING FOR THE BUBBLE SENSOR, THE PATIENT'S BLOOD PRESSURES REACHED AN UNSAFE LEVEL. FOR THE REST OF THE CASE, THE BUBBLE DETECTOR DID NOT GO OFF AGAIN. THERE WERE NO ADVERSE PATIENT EFFECTS AND THE PATIENT RECOVERED IN A NORMAL FASHION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BUBBLE DETECTOR CABLE - ADULT | ACCESSORY | DWA | MEDTRONIC PERFUSION SYSTEMS | BD38 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Life Threatening |