FDA Adverse Event Malfunction Summary report: N

SYNCHRON® LX® I 725 CLINICAL SYSTEM

MDR report key: 1883641 · Received October 26, 2010

Report

Report Number
2050012-2010-01066
Event Type
Malfunction
Date Received
October 26, 2010
Date of Event
October 6, 2010
Report Date
October 26, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K023049
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRIOR TO AND AFTER THE EVENT, ISE QC RESULTS WERE WITHIN THE LAB'S ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT. THE FSE FLUSHED THE MC (MODULAR CHEMISTRY) SAMPLE PROBE AND CHECKED THE ALIGNMENT. THE MC SAMPLE SYRINGE WAS CLEANED AND THE PLUNGER WAS REPLACED. THE FSE ALSO CLEANED THE FLOW CELL AND A SMALL PIECE OF RED RUBBER WAS FOUND ON THE EIC (ELECTROLYTE INJECTION CUP) LEFT VALVE FLAPPER. THE EIC CUP AND VALVES WERE ALSO CLEANED. TROUBLESHOOTING AND INVESTIGATION IS ONGOING.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO OBTAINING ERRONEOUSLY HIGH SODIUM (NA) AND CHLORIDE (CL) RESULTS THAT WERE GENERATED BY THE SYNCHRON LX I 725 CLINICAL SYSTEM. NO ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE SAMPLES WERE REPEATED ON AN ALTERNATE INSTRUMENT AND ON THE ORIGINAL INSTRUMENT, PRODUCING LOWER RESULTS, WHICH WERE REPORTED OUT. THERE WAS NO AFFECT TO PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON® LX® I 725 CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. LX®I 725 N/A

Patients

Seq Age Sex Outcome Treatment
1