FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 1883436 · Received October 21, 2010

Report

Report Number
1824206-2010-10598
Event Type
Malfunction
Date Received
October 21, 2010
Date of Event
September 27, 2010
Report Date
September 27, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE HEAD HI/LOW CYLINDER TO REPAIR THE STRETCHER.

Description of Event or Problem · 1

INFO RECEIVED INDICATES, THE HEAD HI/LOW CYLINDER IS DRIFTING DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 8000

Patients

Seq Age Sex Outcome Treatment
1