CELL-DYN 1700 ANALYZER
Report
- Report Number
- 2919069-2010-00510
- Event Type
- Malfunction
- Date Received
- October 26, 2010
- Date of Event
- October 14, 2010
- Report Date
- October 14, 2010
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION
- Product Code
- GKZ
- PMA / PMN Number
- K870233
- Removal / Correction Number
- 2919069-3/31/09-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). AN EXPANDED INVESTIGATION THROUGH CORRECTION AND REMOVAL FA27SEP2010 WAS CONDUCTED TO EVALUATE THIS ISSUE. THE ROOT CAUSE IDENTIFIED FOR THIS ISSUE IS INSTALLATION OF THE INCORRECT FUSE BY EITHER THE CUSTOMER OR THE FIELD SERVICE REPRESENTATIVE. A PRODUCT CORRECTION LETTER WAS SENT TO THE CUSTOMERS ALONG WITH THE FUSE LABEL TO BE AFFIXED TO THE BACK OF THE INSTRUMENT INSTRUCTING THE CUSTOMER TO CHECK IF THE FUSE MATCHES THE VOLTAGE OF OPERATION AND TO INSTALL THE CORRECT FUSE (PROVIDED IN THE ACCESSORY KIT SHIPPED WITH THE ANALYZER) FOLLOWING REPLACEMENT INSTRUCTIONS PROVIDED IN THE LETTER. THE CUSTOMERS ARE ALSO INSTRUCTED TO ORDER THE CORRECT FUSE IF IT IS NOT PROVIDED IN THE ACCESSORY KIT.
THE CUSTOMER ORDERED THE CORRECT FUSE PER FA27SEP2010 LETTER TO BE INSTALLED ON THE CELL-DYN 1700 ANALYZER. NO IMPACT TO PATIENT RESULTS, PATIENT MANAGEMENT OR USER SAFETY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN 1700 ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | POWER SUPPLY FUSE |