LIFEPAK 500 AUTOMATED EXTERNAL DEFIBRILLATOR
MDR report key: 1883360
·
Received October 21, 2010
Report
- Report Number
- 3015876-2010-01158
- Event Type
- Malfunction
- Date Received
- October 21, 2010
- Date of Event
- September 21, 2010
- Report Date
- September 21, 2010
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K955854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE MAIN PCB ASSEMBLY WAS REPLACED AND THEN PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER. PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS A CRACKED AND OPEN RESISTOR, DESIGNATOR R4.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE FAILED TO DETECT THE ELECTRODES AND THEREFORE, IT WOULD NOT DELIVER THERAPY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.