FDA Adverse Event Malfunction Summary report: N

ARIA RADIATION ONC MGT SYS

MDR report key: 1883350 · Received October 21, 2010

Report

Report Number
2916710-2010-00121
Event Type
Malfunction
Date Received
October 21, 2010
Date of Event
August 31, 2010
Report Date
September 2, 2010
Manufacturer
VARIAN MEDICAL SYSTEMS
Product Code
IYE
PMA / PMN Number
K093527
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BASED ON THE INITIAL REPORT, VARIAN'S MEDICAL ADVISOR HAS IDENTIFIED THAT 200 CGY PER DAY TO A TOTAL OF 5000 CGY FOR 3-DAYS. THE WEDGE WAS LEFT OUT ON THE AP FIELD SO THERE MAY HAVE BEEN A SLIGHT DIFFERENCE IN DOSE DISTRIBUTION AND THEREFORE AN OVERDOSE CORRESPONDING TO WHERE THE THICKER PART OF THE WEDGE SHOULD HAVE BEEN. THIS WAS 60 CGY PER DAY OUT OF 200 CGY AND THE MAXIMUM OVERDOSE WOULD BE 3.6% PER OF THE TOTAL DOSE. THIS IS NOT BELIEVED TO BE A SIGNIFICANT RISK TO PT. THE AVAILABLE INFORMATION SUGGESTS A MALFUNCTION IN THE DEVICE WHICH, SHOULD IT RECUR, COULD POTENTIALLY CAUSE A SERIOUS INJURY. ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT A PLAN WAS IMPORTED AND FIELD A WAS SUPPOSED TO HAVE A WEDGE ATTACHED TO IT, HOWEVER, THEY JUST REALIZED, AFTER 3 FRACTIONS (FIELD A = 107 MU & 60 CGY), THAT THIS FIELD DOES NOT HAVE A WEDGE. EXTRA PLAN DATA: TOTAL DOSE = 5000 CGY. DOSE PER FRACTION = 200 CGY. TOTAL FRACTIONATION = 25 FRACTIONS. CUSTOMER STATES THAT NO ERROR MESSAGE WAS SEEN WHEN IMPORTING IN THE PLAN. DUE TO THE MISSING WEDGE, THERE WAS VARIANCE IN THE PT'S RADIATION PRESCRIPTION. NOTE: CUSTOMER IS AWARE THAT THIS COULD HAVE BEEN USE ERROR. HE WILL BRING THIS TO THE DOCTOR'S ATTENTION TO EVALUATE BEFORE THE 4TH TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARIA RADIATION ONC MGT SYS ACCELERATOR, LINEAR, MEDICAL IYE VARIAN MEDICAL SYSTEMS HRO2770 - ARIARO

Patients

Seq Age Sex Outcome Treatment
1