HEARTMATE II SYSTEM CONTROLLER
Report
- Report Number
- 2916596-2010-00260
- Event Type
- Malfunction
- Date Received
- October 20, 2010
- Date of Event
- June 24, 2010
- Report Date
- September 30, 2010
- Manufacturer
- THORATEC CORP.
- Product Code
- DSQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
THE SYSTEM CONTROLLER WAS RETURNED TO THE MANUFACTURER FOR EVAL AND THE REPORTED EVENT WAS CONFIRMED. THE SYSTEM CONTROLLER WAS CONNECTED TO A MOCK LOOP SET UP WITH A TEST PUMP AND THE SYSTEM WOULD NOT RUN. THE SYSTEM CONTROLLER WAS OPENED AND UPON VISUAL INSPECTION OF THE INTERNAL PRINTED CIRCUIT BOARD (PCB), IT WAS DISCOVERED THAT THE COMPONENT THAT MONITORS THE MOTOR CURRENT WAS NON-FUNCTIONAL. THE OUTPUT FROM THIS COMPONENT WHICH PROVIDES AN INDICATION TO THE MICROPROCESSOR THAT THE PUMP IS CONNECTED DID NOT HAVE A SIGNAL. FURTHER EVAL OF THE DEFECTIVE COMPONENT UNDER X-RAY DID NOT REVEAL ANY DISRUPTIONS ALONG THE CONTACT LINES TO EACH PIN. THE ROOT CAUSE FOR THIS COMPONENT TO BECOME NON-FUNCTIONAL COULD NOT BE CONCLUSIVELY DETERMINED. A REVIEW OF DEVICE HISTORY RECORDS SHOWED NO DEVIATIONS FROM MANUFACTURING OR QA SPECIFICATIONS THAT WOULD AFFECT PUMP FUNCTION OR PERFORMANCE. NO FURTHER INFO IS AVAILABLE. THE MANUFACTURER IS CLOSING ITS FILE ON THIS EVENT.
THE PT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). THE VAD COORDINATOR REPORTED THAT APPROXIMATELY ONE WEEK POST IMPLANT AND WHILE THE PT WAS STILL IN THE HOSPITAL, THE SYSTEM CONTROLLER WAS PRODUCING AUDIBLE ALARMS AND DISPLAYING A VOLTAGE LEVEL OF 0.7 VOLTS WHILE THE PT WAS CONNECTED TO THE POWER BASE UNIT (PBU) OR TO THE POWER MODULE (PM).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTMATE II SYSTEM CONTROLLER | LVAD SYSTEM CONTROLLER | DSQ | THORATEC CORP. | 103696 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Other |