FDA Adverse Event Malfunction Summary report: N

EON IPG

MDR report key: 1883213 · Received October 20, 2010

Report

Report Number
1627487-2010-03001
Event Type
Malfunction
Date Received
October 20, 2010
Date of Event
September 21, 2010
Report Date
September 21, 2010
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION DIVISION
Product Code
LGW
PMA / PMN Number
P010032
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DEVICE HISTORY RECORD FOUND A NONCONFORMANCE; HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE PT RECEIVED HER SCS SYSTEM ON (B)(6) 2005. IT WAS REPORTED THAT THE PT LOST STIMULATION ABOUT THREE YEARS AGO FOLLOWING A FALL. THE IPG WAS REPLACED ON (B)(6) 2010 AND EFFECTIVE STIMULATION WAS RECAPTURED. THE EXPLANTED DEVICE WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. NO FURTHER ISSUES HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EON IPG TOTALLY IMPLANTABLE PULSE GENERATOR LGW ST. JUDE MEDICAL - NEUROMODULATION DIVISION 3716 46130A

Patients

Seq Age Sex Outcome Treatment
1